Specifications
Dimensions
Size (W x H x L): 36.8cm x 24.6cm x 33.6cm (14.5in x 9.75in x 13.25in)
Weight: 15.9 kgs (35 lbs.)
Power Ratings
Power Requirements:
110 VAC range (100 to 127VAC), 60 Hz
220 VAC range (220 to 240VAC), 50 Hz
External Voltage Selector Switch
Power Consumption:
Maximum: 630 kW hours per year
Nominal: 315 kW hours per year
Battery Life:
75-80 ampere-hour 12VDC deep-cycle gel-cell battery: 20 hours
35-40 ampere-hour 12VDC deep-cycle gel-cell battery: 10 hours
Internal battery: 1 hour
Physical Characteristics and Display
Front Panel Touch Pads:
Alarm Silence/Reset: Push button to silence alarm or to reset; used with battery test button to read operating hours on patient pressure meter
Battery Test: Push button to show charge level on lower window of patient pressure meter
Front Panel Controls
Mode: Standby, Assist/Control, SIMV or Pressure Cycle
Pressure Limit: Locking knob sets presure from 15 to 50 cm H2O/hPa, or disables limit feature
Low Alarm: Set kimit for low pressure alarm (2-32 cm H20/hPa in increments of 2 cm H20/hPa)
High Alarm Limit: Set limit for high pressure alarm (Assist/Control, SIMV modes) or limit for pressure cycle mode (15-90 cm H20/hPa in increments of 5cm H20/hPa)
Volume: Push-to-turn knob (continuously adjustable from 100 to 2200 ml)
Breath Rate: Sets breating rate (1-20 BPM in increments of 1 BPM and 22-38 BPM in increments of 2 BPM)
Inspiratory Time: Sets time from 0.5 to 1.0 seconds in increments of 0.1 seconds, 1.2 seconds, and 1.5-5.5 second in increments of 0.5 seconds
Breathing Effort: Push-to-turn knob sets pressure level to detect breathing effort (-10 cm H20/hPa to +10 cm H20/hPa)
Front Panel Input
Patient Pressure: Port for connection to proximal pressure line of patient circuit
Front Panel Output
Patient Air: 22 mm x 15 mm tube for connection to bacteria filter
Exhalation Valve: Port for connection to exhalation valve of patient circuit
Indicators, Normal Events
Patient Pressure: Displays patient pressure -10 to +100 cm H20/hPa; can also display battery charge and machine hours of operation
Breathing Effort: Green LED activated by adequate patient breathing effort
Power: LEDs indicate operating power source (green AC Pwr/Batt Chg, amber External Battery, flashing amber Internal Battery)
Alarms: Flashing red LEDs (low pressure/apnea, low power, high pressure, setting error, power switchover
Rear Panel Controls
AC Circuit Breaker: Internal to AC power switch (1 for On, 0 for Off)
Rear Panel Inputs
Inlet Filter: Intake for patient air, screw-off cap for filter change
External Battery: Connection for 12 VDC battery
Communications Port: Calibration information during service procedures
Rear Panel Outputs
Vent: Cooling vent for internal ventilator components
Remote Alarm: Connection for optional alarm accessories
Communication Port: Connection for optional printer and digital display
Alarms
Audible Alarms
Pulsating Tone: Low Pressure/Apnea; low power; high pressure; setting erroy; power switchover
Steady Tone: Loss of microprocessor control
Reminder Tone: Every 5 minutes when powered by internal battery and each time accessory printer generates report
Ventilator Ratings
Resistance Factor: Less than 0.05 cm H2O/hPa/lpm
Emergency Pressure Relief: 100 cm H2O/hPa
Flow: 20 to 100 lpm
Environmental
Operating Temperature: 5°C to 40°C (41°F to 104°F
Storage Temperature: -15°C to 50°C (5°F to 122°F)
90% RH non-condensing, allow at least 1 hour temperature stabilization before use when moving from non-operating to operating
Other
Preventive Maintenance: By qualified professional every 6000 hours or recertification every 12 months, whichever is first