Specifications
Dimensions
Size (W x H x D):
VS System: 33.02cm x 27.94cm x 36.83cm (13.0in x 11.0in x 14.5in)
VSO2 System: 33.02cm x 27.94cm x 36.83cm (13.0in x 11.0in x 14.5in)
Docking Bay:
VS System: 30.48cm x 2.54cm x 27.94cm (12.0in x 3.0in x 13.5in)
VSO2 System: 30.48cm x 2.54cm x 27.94cm (12.0in x 3.0in x 13.5in)
External Battery:
VS System: 2.54 cm (1.0 in)
VSO2 System: 2.54 cm (1.0 in)
Weight:
VS System: 15 kg (33 lbs.)
VSO2 System: 15.5 kg (34 lbs.)
Docking Bay
VS System: 1.1 kg (2.5 lbs.)
VSO2 System: 1.1 kg (2.5 lbs.)
External Battery
VS System: 4.5 kg (10 lbs.)
VSO2 System: 4.5 kg (10 lbs.)
Construction: Internal impact resistant frame, durable rubber exoskeleton
Ventilator Parameters
Controls
Remote Alarm (Optional)
Ranges: ON/OFF
Tolerance: N/A
Inspiratory Hold
Ranges: 6 second max.
Tolerance: Greater of ±2 cmH2O or 10%
100% O2 (VSO2 only)
Ranges: ON/OFF (3 min. max)
Tolerance: +0%, -5%
Flow Cal
Ranges: ON/OFF
Tolerance: N/A
Over Pressure Relief
Ranges: 20 to 130 cmH2)
Tolerance: ± cmH2O
Alarms
High Pressure Alarm Limit
Ranges: 5 to 120 cmH2O
Tolerance: Setting of 5 to 20 cmH2O ± 2 cmH2O; Setting of 21 to 120 cmH2O ±4 cm H2O
Low Pressure Alarm Limit
Ranges: OFF, 2 to 60 cmH2O
Tolerance: Setting of 2 to 20 cmH2O ± 2 cmH2O; Setting of 21 to 60 cmH2O ±4 cm H2O
Low Minute Volume Alarm
Ranges: 0.1 to 99.9 L
Tolerance: Greater of: ±15% or 30 ml
Apnea Interval
Ranges: 10 to 60 sec.
Tolerance: ±0.5 sec.
Backup Breath
Ranges: Greater of: 12 bpm or set breath rate
Tolerance: Greater of: ±1 breath or 10% of breath period
Low O2 (VSO2 only)
Ranges: 35 psig
Tolerance:
High O2 (VSO2 only)
Ranges: 65 psig
Tolerance:
Ventilator Inoperative
Ranges: Red indicator
Tolerance: N/A
Alarm Silence
Ranges: 60 sec. max.
Tolerance: ±l second
Alarm Volume
Ranges: 50 to 70 dBa
Tolerance: ± 5 dBa
Indicators
Patient Effort: Yellow Indicator
Power Indicators
On : Green indicator
AC : Green indicator
Internal Battery Battery Power: Green, Yellow, Red indicator
Internal Battery Charge Status: Green, Yellow, Red indicator
External Battery Battery Power: Green, Yellow indicator
External Battery Charge Status: Green, Yellow, Red indicator
Pneumatic Specifications
Oxygen Inlet:
VS Oxygen Bleed In: Max. 80 lpm or 0.5 PSI (.035 Bar)
VSO2 Pressurized Oxygen: 40 to 60 PSI (2.8 to 4.2 Bar)
Environmental
Operating Temperature: 5°C to 40°C
Storage Temperature: -20°C to 60°C
Relative Humidity:
Operating: 15 to 95% relative (non-condensing)
Storage: 10 to 95% relative (non-condensing)
Sound Level: < 50 dBa (rms) at one meter
Shock: Complies with IEC 68-2-27 (MIL-STD-810E)