Dimensions
Size (W x H x D): 254mm x 88.9mm x 139.7mm (10in x 3.5in x 5.5in)
Weight: 2.2 kg (4.8 lbs.)
Power Ratings
Battery Life: 2 hours use on 4 hour full charge
Battery Type: Nickel Cadmium, 6VDC, internal rechargeable but not user replaceable
AC Charge Power Requirements: 115 VAC, 60 Hz
Displays
Model Number: 9004
Display: Vacuum Flourescent
Displayed Trends: Graphical display, factory selectable sample rate, user selectable trend
Displayed: 16 hour display at 12 second sample rate
Waveforms: 10 hour display at 4 second sample rate
Selectable COM2 or Co2/Plethysmogram waveform
| Alarms
Physiological: Dual pitch, adjustable volume. User-selectable high and low limits for all displayed parameters. Screen indicator sfor each parameter
System: Single pitch, adjustable volume
Silence: Temporary 2 minute or OFF LED indicator
CO2
Measurement: Non-Dispersive IR absorption
Measurement Range: 0 to 10% CO2
Units: User selectable: mm Hg, Vol, %, kPa
Display Range: 0-100 mmHg or 0 to 10.0% Co2 or 0 to 13.3 kPa
Accuracy: ±2 mmHg; ±0.3%; ±0.3 kPa, or 4% of reading, whichever is greater (0 to 10% CO2)
Resolution: 1 mmHg; 0.1%; 0.1 kPa
Calibration: 2 point Cal
N2O Compensation: User selectable 40% (default = OFF)
Response Time: 375 msec to 90% of value
Averaging: 4-breath average
Respiration Rate
Range: 0 to 150 bpm
Accuracy: ±1 bpm
Averaging: 4-breath average
Flow Rates
Flow Rates: 150 ml/min ±20
|
SpO2 (Optional)
Range: 0 to 100%
Accuracy: ±2% at 70-100%, ±3% at 50-69%
Pulse Rate (Optional)
Range: 30 to 254 bpm
Accuracy: ±2% at 30 to 250 bpm
FiO2 (Optional)
Range: 0 to 100%
Accuracy: ±2%
Response Time: < 10 sec. to 90%
Drift: < 2% full scale per month
Standards
Safety Approvals: EN60601-1, EN60601-1-1, UL 544, CSA Std. C22.2 No. 125-1984
EMC: EN60601-1-2
|